Assumptions for Parameter Estimations
- TR >> T2* (T2 in the absence of refocusing echoes).
- A monoexponential relaxation process governs T2 decay and T1 recovery.
- The rf pulse flip angles are uniform through the slice thickness.
- The rf pulse duration is negligible.
- Inhomogeneity of the non-selective 180 rf pulse is assumed to be negligible.
- Partial volume effects, image data set misregistration, flow, self-diffusion, chemical shift, susceptibility effects at tissue interfaces are negligible.
The parameter estimates influenced by incorrect assumptions may have poor figure-of-merit statistics associated with the estimates, or the estimation may simply fail.
All equations are from Stark & Bradley. Magnetic Resonance Imaging. Moseby Year Book, St. Louis, 1992, unless otherwise indicated.
|